A1 – SafeTech

Weighing safety cabinets: the science behind safe and accurate weighing

Why a professional scale is much more than just a “cabinet” with a HEPA filter

The science behind safe, vibration-free and accurate weighing environments

At first glance, an analytical balance appears to be a relatively simple construction: a transparent housing, a fan and a HEPA filter. But anyone who works with analytical balances, microbalances or potentially hazardous powders knows that reality is considerably more complex.

A well-designed weighing scale must combine two objectives that can influence each other: maximum protection for the user and a stable, accurate weighing environment.

At A1-SafeTech, this challenge forms the basis for the design of specialized weighing safety cabinets. By paying attention to aerodynamics, vibration control, filtration, and ergonomics, solutions are created that go beyond conventional extraction systems.

1. Air flow: the invisible factor in accurate road measurements

When weighing small amounts of powder, even slight air movements can affect the measurement results.

Analytical balances and microbalances are particularly sensitive to changes in airflow. Turbulence can cause forces on the weighing platform, the weighing container, or the material itself. This can cause the displayed mass to fluctuate and take longer to reach a stable measurement value.

High air velocity does not automatically mean better protection either. It mainly depends on how the air is directed through the work opening and along the work area.

The weighing safety cabinets from A1-SafeTech are designed with an optimized, turbulence-free airflow concept. This helps to combine controlled airflow with the stability necessary for precise weighing processes.

Not only the amount of air drawn in is important, but especially how this air moves through the weighing environment.

2. Why a conventional extractor hood is not always suitable

A conventional laboratory acid cabinet is usually designed to safely remove vapours and gases. A specialised powder weighing cabinet must also take into account the properties of solid particles and the sensitivity of analytical weighing equipment.

In the case of unfavorable airflow, turbulence can develop around the balance, the operator's hands, and the materials used.

This can lead to:

  • Unstable weighing values and longer stabilization times.
  • Unwanted dispersion of fine powder particles.
  • Disruption of the weighing process due to air movements.
  • An increased risk of contamination of the work area.
  • Unnecessary burden on the operator.

A specialized weighing safety cabinet is therefore not simply a smaller version of a conventional acid cabinet.

3. Vibrations: a problem often underestimated

In addition to airflow, mechanical vibrations are an important factor in accurate road construction.

Fans and electric motors can produce vibrations that are transmitted to the balance via the construction of a weighing platform. Especially in microbalances, where very small mass differences are measured, this can have an impact on the stability and reproducibility of the measurement.

In the A1-SafeTech ST1 series, the fan and filter unit are located separately from the transparent weighing housing.

Because no fan motor is mounted directly on the housing, an important potential source of mechanical vibrations is eliminated.

Combined with a suitable, stable balance setup, this contributes to a quiet and reproducible weighing environment.

4. HEPA filtration: important, but not the whole story

HEPA filtration plays an essential role in capturing fine particles.

Depending on the configuration, A1-SafeTech systems can be equipped with high-quality H14 HEPA filtration, including a filter design with dual filter layers.

But the presence of a HEPA filter alone does not yet provide sufficient information about the protection that a complete air purifier offers.

The actual performance is determined in part by:

  • The airflow at the work opening.
  • The design of the housing.
  • The position and operation of the extraction fan.
  • The sealing and integrity of the filter system.
  • The operator's method of operation.
  • The correct installation, inspection and maintenance.

Therefore, a good weighing safety cabinet must be assessed as a complete system.

5. Containment: protection must be demonstrably achieved

When working with active pharmaceutical ingredients (API’s), toxic substances and other potentially dangerous powders, operator protection is central.

The purpose of containment is to prevent harmful particles from entering the user’s breathing zone during weighing or processing.

Concepts such as Occupational Exposure Limit (OEL), Occupational Exposure Band (OEB) and containment performance play an important role in this.

A1-SafeTech offers solutions for applications with high containment requirements, including configurations for OEB 5 applications.

The suitability of a specific weighing cabinet, however, depends on the application, the materials used, the configuration, and the demonstrated containment performance.

Where necessary, containment measurements can be used to assess performance under established test conditions.

A safe weighing environment is not determined solely by the specifications of the components, but by the performance of the entire system.

6. Ergonomics and accessibility

Safety and measurement accuracy are important, but a weighing scale should also be comfortable and practical in everyday use.

An unfavorable working posture, limited mobility, or poor visibility when balancing can complicate the weighing process.

The transparent PMMA housings of A1-SafeTech provide good visibility into the work area. The design takes into account accessibility, working posture, and the operation of analytical balances.

Depending on the model and configuration, additional ergonomic features may be available.

This allows safety, accuracy and ease of use to be combined.

7. Installation, qualification and periodic inspection

Even a carefully designed weighing scale may not function properly when it is installed or used incorrectly.

The location in the laboratory, environmental air flows, the setup of the balance and the work procedures can all have an impact on performance.

Therefore, proper installation and appropriate qualification are essential.

Depending on the application, the following tasks can be performed, among others:

  • Inspection of the installation and the technical configuration.
  • Measurement of airflow and inlet speed.
  • Checking the HEPA filter system.
  • Functional control of alarms and security systems.
  • Documentation for IQ/OQ.
  • Additional containment measurements when required.

Periodic inspection, maintenance and re-qualification also contribute to maintaining the desired performance.

8. A1-SafeTech and Prolyse: specialized solutions for the Benelux

Prolyse B.V. represents A1-SafeTech exclusively in the Benelux region.

We advise laboratories when choosing suitable weighing safety cabinets and containment solutions, taking into account not only the dimensions or the desired extraction capacity, but especially the actual application.

What balance is used? With which substances is work being carried out? What precision is required? What containment objective must be achieved?

By answering these questions in advance, a solution can be selected that meets the laboratory's technical and safety requirements.

Prolyse also provides coordination of delivery, installation, qualification and service, in collaboration with the specialized technicians of A1-SafeTech.

Want to know more about safe and accurate weighing?

Are you looking for a weighing safety cabinet for analytical balances, microbalances, or the processing of potentially dangerous powders?

Contact Prolyse for advice on the possibilities of A1-SafeTech. We are happy to help you select a suitable solution for your laboratory.

View the A1-SafeTech weighing safety cabinets or request free technical advice.

Contact with Prolyse

Would you like to receive a free quote or have any questions? Please contact us via:

Phone: +31 (0)162 383100 or click here for the contact form.

+31 (0)162 383100
A1-SafeTech ST1 weighing safety cabinet for analytical balances

About Prolyse

Prolyse BV is a company located in Heerle, North Brabant, the Netherlands, close to the Belgian border. We were founded in 2009 and work closely with agents, dealers, freelancers and suppliers. We often work with a flexible team.

Services

We provide our services in analytical chemistry, process industry, pharmaceutical and life science laboratories.

Read more

Support

We offer a high level of service and support that you can expect, even after the warranty period!

Read more

Satisfied customers

A recommendation from a satisfied customer is absolutely the best reference; it is our absolute conviction!!” We have customers in the BeNeLUX countries, Ireland, Germany, the Czech Republic and even in Asia.

Read more

Do you have a project?
Let's talk & Making a success of your project

We are ready to help you. Our expert is there, just send a message.