Sampling Cannulas are used to take samples from the dissolution vessels. Choosing a cannula is important because the sampling method must not have a negative effect on the accuracy of the dissolution – and must not give incorrect results.
Key points of the sampling cannulas:
- Sampling position: this is very important because the concentration of the solution can vary at different heights. It has been agreed that the sampling should take place halfway up the paddle and the liquid level in the dissolution vessel. Additionally, this should be done at least 10mm from the vessel wall.
- Cannula Length: These vary in length from 120 to 380mm. This depends on the contents of the vessel, e.g. 500ml or 900ml.
- Cannula Material: These are usually made of stainless steel or PEEK so that there is no interference with the active components.
- Stoppers: the cannula can be used with a so-called stopper so that sampling always takes place at the same height.
QLA Guarantee
All QLA Products are manufactured in accordance with ISO9001 standards and are supplied with certificates and authenticity holograms.
Height
Sampling at the correct height is very important. For this purpose you can use a stopper on the cannula.
Filtering
On the cannula you can slide a so-called full flow or finger tip filter, so that during the actual sample collection the filtering process also takes place.
Moving probes
In automated setups, a movable manifold is often used with all measuring and sampling probes attached to it. At the time of sample collection, these are moved downward, a sample is taken, and the probes are returned upward.